- Features
- Specifications
- Downloads
- Enquiry
Product Features
- Compact Plastic Casing, Anti-Shock, Ergonomic, Easy of Operation
- Wide Range of Syringe Brands - 5 ml, 10 ml, 20 ml, 30 ml, 50 ml
- High Water-Proof Performance, Automatic Syringe Type Sensing
- Four kinds of injection modes: Rate mode,Time mode,Body weight mode,Link Mode
- Double CPU Operation, makes injection Safe and Reliable
- Excellent Flow-Rate Accuracy with Precision Stepping Motor
- Audio and Visual alarms - Ensure clinician awareness and prompt intervention
- Intelligent Pressure Monitoring - Precisely detects over running and obstruction
- Accumulated injection volume recordable when "Bolus"function is ON
- Flexible, Power Supply - DC/AC & Built-in Lithium Battery
- Wireless data transmission with infusion supervision system
Syringe Pump – S750i:
Infusion
Syringe Channel – Double channel
Syringe Capacity
5, 10, 20, 30, 50/60 ml syringes
Syringe Selection / Calibration
Pre-Calibrated Syringe setting and type for Standard Brands, Auto-Calibration for other Brands
Operating Mode
Rate Mode, Time Mode, Body Weight Mode
Flow Rate Range
0.1~100 ml/h (0.1 ml/h step) 5 ml syringe, 0.1~200 ml/h (0.1 ml/h step) 10 ml syringe,
0.1~400 ml/h (0.1 ml/h step) 20 ml syringe, 0.1~600 ml/h (0.1 ml/h step) 30 ml syringe
0.1~1500 ml/h (0.1 ml/h step) 50/60 ml syringe
Bolus
Manual / Preset - Auto bolus
Flow Rate Accuracy
Accuracy including syringe: within ± 2%, Mechanical accuracy: within ± 1%
Accumulated Volume
0.1~9999.9ml
KVO Rate
0.0 - 5.0 ml/hr Adjustable, 0.1 ml/hr as Default
Setting of Delivery Limit
0.1~999.9ml
Occlusion Detection Pressure
‘L’:40 to 60 kPa , 'M' : 61 -90 kPa, ‘H’: 91 -120 kPa
Display
Screen
3.0 inch Colour LCD screen for Individual Syringe Channel
Information
‘Syringe Rate’, ‘Volume Limit’, ‘Syringe brand and Size’, ‘AC power / Battery Charge Indicator’, ‘Alarm Information’, Pressure level, ‘Real volume’
Safety Features
Acoustic Alarms
‘Occlusion’, ‘Near Empty’, ‘Infused/KVO’, ‘Syringe Empty’, ‘Plunger disengaged’, ‘Flow rate Error’, ‘Low Battery’, ‘Battery Exhausted’, ‘AC power OFF’, ‘Volume over limited’, ‘System error’, ‘Confirm/back’, 'Auto Pre-set Bolus', ‘Check Syringe Clamp’
External Visual Alarm
Dual-Redundant, Dual-Colour Visual Alarms
Interfaces
Ports
Ambulatory port / 12VDC input, Nurse Call, RS232 Data port
Wireless LAN
Optional
Infusion Management
Wireless Integration with Omya FM1000i model (Infusion Management System)
Power
Power Supply
Integrated Power Supply, rated Voltage AC 100~240V, 50~60 Hz, Power Consumption: 25 VA
External Fuse
3 A / 250 V (User Replaceable)
Battery
Battery life
Operates for 6-8 hrs from Full Charge (5 ml/hr at 30°C under Normal Conditions)
Battery Operation
Advanced Rechargeable Lithium Polymer battery: 7.4 V, 6000mAh
Battery Charge
When AC Power is ON, Battery will automatically Recharge (about 4 hrs to 80% charge)
Safety Specification
Product Classification
According to the MDD order 93/42/EEC Appendix ? Classification Rule 11: class?b
Standards
EN / IEC 60601-1 and EN / IEC 60601-2-24 / CE Marking in Compliance with the 93/42/EEC European Medical Device Directive
Electric Shock Protection/Class/Level
Class II, Equipment type BF, Continuous Operation
Ingress Protection
IP 44 (Complete protection against ingress of small solid foreign bodies, and Water splashed against the Device from any direction)
Work Operation
Continuous
Environment
Operation Conditions
Ambient temperature: + 5 to + 40°C, Atmosphere Pressure: 86 ~106 kPa, Relative humidity: < 80%
Storage Conditions
Ambient temperature: – 20 to +55°C, Atmosphere Pressure: 50 ~106 kPa, Relative humidity: < 95%
Non-corrosive gas and well-ventilated room
Physical Properties
Dimension
260mm (L) x 137.5mm (W) x 245mm (H)
Weight
3.7 kg (with clamp base)